Troubleshooting
6-9
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0105003
Warning
Filament is over
current by SW
monitor
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1. Check DC power supplier
and fuse F5 on the Mother
board (backplane).
2. Check cables and
connections
3. Check the MCB.
4. Check the FDB.
5. Replace the monoblock.
0106001
Warning
6954 chip
verification is
failed: The
written value is
not same with
the read back
value
Control Panel exception.
Exposure is disabled. Please
restart the system. If this issue
persists, please call for service.
1. MAX6954 chip of Display
Board may be abnormal
2. Maybe connection issue
in Cable between Display
Board and Mother board
3. Power supply may be
abnormal
4. MCB SPI channel may be
abnormal
0106002
Warning
7301 chip
verification is
failed: The
written value is
not same with
the read back
value
Control Panel exception.
Exposure is disabled. Please
restart the system. If this issue
persists, please call for service.
1. 7301 chip of Display
Board may be abnormal.
2. Maybe connection issue
in Cable between DB and
Mothor board
3. Power supply may be
abnormal
4. Maybe MCB SPI channel
may be abnormal.
0106003
Warning
6954 channel
of SPI bus TX is
time out
Control Panel exception.
Exposure is disabled. Please
restart the system. If this issue
persists, please call for service.
1. MCB may be abnormal
2. Motherboard and Display
Board cable connection
may be abnormal
3. DB may be abnormal
0106004
Warning
6954 channel
of SPI bus RX is
time out
Control Panel exception.
Exposure is disabled. Please
restart the system. If this issue
persists, please call for service.
MCB SPI write to Display
Board (Nixie tube
) time out:
1. MCB may be abnormal
2. Motherboard and Display
Board cable connection
may be abnormal
3. DB may be abnormal
0106005
Warning
7301 channel
of SPI bus TX is
time out
Control Panel exception.
Exposure is disabled. Please
restart the system. If this issue
persists, please call for service.
MCB SPI write to Display
Board (Nixie tube
) time out:
1. MCB may be abnormal
2. Motherboard and Display
Board cable connection
may be abnormal