Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-10
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
3. DB may be abnormal
0106006
Warning
7301 channel
of SPI bus RX is
time out
Control Panel exception.
Exposure is disabled. Please
restart the system. If this issue
persists, please call for service.
MCB SPI writes to Display
Board (Function LED) time
out
1. MCB may be abnormal
2. Motherboard and Display
Board cable connetction
may be abnormal
3. DB may be abnormal
0107001
Warning
Left hand
switch is stuck
since safe line
error
Left hand switch stuck.
Exposure is disabled. Please
check the left handswitch and
restart the system. If this issue
persists, please call for service.
N/A
0107002
Warning
Right hand
switch is stuck
since safe line
error
Right hand switch stuck.
Exposure is disabled. Please
check the right hand switch
and restart the system. If this
issue persists, please call for
service.
N/A
0107800
Warning
During bootup ,
left hand switch
is pressed
Left hand switch pressed
during boot up. Please release
the left hand switch.
N/A
0107801
Warning
During bootup ,
right hand
switch is
pressed
Right hand switch pressed
during boot up. Please release
the right hand switch.
N/A
0108001
Warning
Left foot switch
is stuck since
safe line error
Left foot switch stuck.
Exposure is disabled. Please
check the left foot switch and
restart the system. If this issue
persists, please call for service.
N/A
0108002
Warning
Right foot
switch is stuck
since safe line
error
Right foot switch stuck.
Exposure is disabled. Please
check the left foot switch and
restart the system. If this issue
persists, please call for service.
N/A
0108800
Warning
During bootup ,
left foot switch
is pressed
Left foot switch pressed during
boot up. Please release the left
foot switch.
N/A