Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-10
ID
Warning
Description
(Displayed in User Interface)
of SPI bus RX is
time out
Exposure is disabled. Please
restart the system. If this issue
persists, please call for service.
MCB SPI writes to Display
Board (Function LED) time
out
1. MCB may be abnormal
2. Motherboard and Display
Board cable connetction
may be abnormal
switch is stuck
since safe line
error
Exposure is disabled. Please
check the left handswitch and
restart the system. If this issue
persists, please call for service.
switch is stuck
since safe line
error
Exposure is disabled. Please
check the right hand switch
and restart the system. If this
issue persists, please call for
left hand switch
is pressed
during boot up. Please release
the left hand switch.
right hand
switch is
Right hand switch pressed
during boot up. Please release
the right hand switch.
is stuck since
safe line error
Exposure is disabled. Please
check the left foot switch and
restart the system. If this issue
persists, please call for service.
switch is stuck
since safe line
error
Exposure is disabled. Please
check the left foot switch and
restart the system. If this issue
persists, please call for service.
left foot switch
is pressed
Left foot switch pressed during
boot up. Please release the left
foot switch.