Troubleshooting
6-11
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0108801
Warning
During bootup ,
right foot
switch is
pressed
Right foot switch pressed
during boot up. Please release
the right foot switch.
N/A
0109800
Warning
Key switch is
rotated to "X-
ray Off" position
Key Switch in Xray Off.
Exposure is disabled. To enable
exposure, please turn it to On.
N/A
0109801
Warning
Key switch is
rotated to "Off"
position
Keyswitch in Off. Exposure and
lift column motion are
disabled. To enable exposure
or lift column motion, please
turn it to On.
N/A
0110800
Warning
Emergency
switch is
pressed
Emergency switch pressed.
Exposure and lift column
motion are disabled. To enable
exposure or lift column motion,
please release the switch.
N/A
0112500
Warning
DAP event
buffer overflow
Mainframe buffer overflow. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
1. Don't follow calibration
procedure to perform
calibration. (Reboot system
and Login Serivice Interface,
directly perform collimator
calibration to resolve this
problem.)
2. MCB may be abnormal
0112501
Warning
DAP parameter
buffer overflow
Mainframe buffer overflow. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
1. Don't follow calibration
procedure to perform
calibration. (Reboot system
and Login Service Interface,
directly perform collimator
calibration to resolve this
problem.)
2. MCB may be abnormal
0112800
Warning
DAP is not
calibrated
without filter
Dose not calibrated for the
tube without filter option.
Please call service to calibrate
dose.
0112801
Warning
DAP is not
calibrated with
filter
Dose not calibrated for the
tube with filter option. Please
call service to calibrate dose.
0113002
Warning
WKS UART
write fault
Communication exception. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
MCB serial port writes to
WKS fault.
1. MCB may be abnormal.
2. Motherboard may be