Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-12
ID
Warning
Description
(Displayed in User Interface)
3. Cable connection may be
abnormal
comservo
protocol is
incompatible
with WKS
comservo
exception. The mainframe
stopped. Please call for service.
1. WKS and MF version may
be not compatible.
Communication exception. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
between WKS and MF
failure:
1.Bundle cable may be
abnormal.
2. Cable connection
between WKS computer
serial port and
motherboard may be
abnormal.
3. WKS computer serial port
buffer overflow
Communication exception. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
from MF to
WKS time out
Communication exception. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
1. check interconnect cable
2. check MCB input power
buffer overflow
Communication exception. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
finds WKS in 4
minutes after
system boot up
Communication exception. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
1. check interconnect cable
2. Check PDU output
voltage.
3. Check Workstation
mainboard.
Can’t detect the workstation.
The mainframe stopped.
Please restart the system. If
this issue persists, please call
1. check interconnect cable
2. Check PDU output
voltage.
3. Check Workstation