Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-12
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
abnormal.
3. Cable connection may be
abnormal
0113500
Warning
Mainframe
comservo
protocol is
incompatible
with WKS
comservo
protocol
Communication protocol
exception. The mainframe
stopped. Please call for service.
1. WKS and MF version may
be not compatible.
0113501
Warning
RCC time out
Communication exception. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
Serial communication
between WKS and MF
failure:
1.Bundle cable may be
abnormal.
2. Cable connection
between WKS computer
serial port and
motherboard may be
abnormal.
3. WKS computer serial port
may be abnormal.
0113502
Warning
WKS message
buffer overflow
Communication exception. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
1.MCB may be abnormal
0113503
Warning
sendMessage
from MF to
WKS time out
Communication exception. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
1. check interconnect cable
2. check MCB input power
0113504
Warning
RCC-UART Tx
buffer overflow
Communication exception. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
1. MCB may be abnormal
0113505
Warning
Mainframe not
finds WKS in 4
minutes after
system boot up
Communication exception. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
1. check interconnect cable
2. Check PDU output
voltage.
3. Check Workstation
mainboard.
4. Check the MCB .
0113506
Warning
Ping time out
Can’t detect the workstation.
The mainframe stopped.
Please restart the system. If
this issue persists, please call
for service.
1. check interconnect cable
2. Check PDU output
voltage.
3. Check Workstation
mainboard.

Table of Contents