Troubleshooting
6-13
ID
Warning
Description
(Displayed in User Interface)
port UART write
fault
Camera control exception. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
port
transmission
Camera control exception. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
port RCC time
out
Camera control exception. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
MCB and camera failure.
1.MCB may be abnormal.
2.Camera may be
abnormal.
3.F7 (fuse) on mother board
(backplane) may be
abnormal.
4.Bundle cable may be
abnormal
5. DC POWER SUPPLY +5V,
+12V AND
+24V(PN:5085801) may be
abnormal.
6. I.I power supply and I.I
port
unexpected
Camera control exception. The
mainframe stopped. Please
restart the system. If this issue
persists, please call for service.
protocol issue
1.Camera may be
fault during
calibration
exception. The mainframe
stopped. Please restart the
system. If this issue persists,
calibration
1.MCB
fault when
saving product
information
exception. The mainframe
stopped. Please restart the
system. If this issue persists,
during MA
calibration
exception. The mainframe
stopped. Please restart the
system. If this issue persists,