Troubleshooting
6-17
ID
Warning
Description
(Displayed in User Interface)
memory
segment failed
restart the system. If this issue
persists, please call for service.
Check if there exists some
unwanted share memory in
the system, or check if
software is installed
open a named
semaphore
failed
Semaphore exception. Please
restart the system. If this issue
persists, please call for service.
Check if there exists some
unwanted semaphore in
the system, or check if
software is installed
semaphore
failed
Semaphore exception. Please
restart the system. If this issue
persists, please call for service.
Check if the semaphore is
valid, or check if the
maximum allowable value
for the semaphore will be
Image FIFO full. The mainframe
stopped. Please restart the
system. If this issue persists,
please call for service.
thread priority. Complied
SW version (Debug or
Release)
configuration
file failed
IP configuration exception.
Image quality may be affected.
Please restart the system. If
this issue persists, please call
Check if IP configuration file
crashed
configuration
file failed
Setup configuration exception.
Some system setting may be
missed. Please restart the
system. If this issue persists,
Check if Setup configuration
file crashed
section failed
Execution exception. Please
restart the system. If this issue
persists, please call for service.
section failed
Execution exception. Please
restart the system. If this issue
persists, please call for service.
section failed
Execution exception. Please
restart the system. If this issue
persists, please call for service.
failed
keyboard display exception.
The display may be abnormal.
Please ensure no key is
pressed on keyboard, then
press “Auto B/C” key and
“Negate” key. If this issue
persists, please call for service.
Check system IO state and
keyboard connection