Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-26
ID
Warning
Description
(Displayed in User Interface)
Exposure config
file
Exposure configuration loading
exception. Some configuration
may be missed. Please restart
the system. If this issue
persists, please call for service.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
summary
modality type
failed
Get Dose summary modality
setting failed. Please restart
the system. If this issue
persists, please call for service.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
for importing
image failed
Import images from CD/DVD
exception. Please try again. If
this issue persists, please call
for service.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
Loading message fail. Please
restart the system. If this issue
persists, please call for service.
module get
OSGI service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
application
exception
exception. Please restart the
system. If this issue persists,
please call for service.
module
unregister OSGI
service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
exception
failure. Please restart the
system. If this issue persists,
please call for service.
XmlFileParseUtil
ity for
ConnectivitySet
up.xml failed
Connectivity configuration
loading failure. Some
configuration may be missed.
Please restart the system. If
this issue persists, please call
for service.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.