Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-28
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0304520
Warning
Init MWL server
list failed
Connectivity configuration
loading failure. Worklist server
list missed. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0304521
Warning
Load anatomic
region
configuration
file failed
Anatomic Regions list missed.
Please restart the system. If
this issue persists, please call
for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
3. If not work, re-install OS.
0304523
Warning
Init date & time
format failed
Setup configuration loading
failure. Date/Time format
setting missed. Please restart
the system. If this issue
persists, please call for service.
0304524
Warning
Get string failed
Loading message fail. Please
restart the system. If this issue
persists, please call for service.
0304525
Warning
Stop
ExamManagem
ent failed
Stop ExamManagement failed.
Please restart the system. If
this issue persists, please call
for service.
0304526
Error
ExamManagem
ent module
register OSGI
service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0304527
Error
ExamManagem
ent module get
OSGI service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0304528
Error
ExamManagem
ent module
unregister OSGI
service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0304529
Warning
Stop
application
exception
Close ExamManagement
exception. Please restart the
system. If this issue persists,
please call for service.