Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-34
ID
Warning
Description
(Displayed in User Interface)
zoom
information
from IE failed
Workstation communication
exception. Can’t get image and
zoom roam data from the left
monitor. Please restart the
system. If this issue persists,
please call for service.
2. If not work, re-install
Workstation software.
failed
Workstation communication
exception. Can’t set zoom
roam data to the left monitor.
Please restart the system. If
this issue persists, please call
for service.
2. If not work, re-install
Workstation software.
module register
servant
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
2. If not work, re-install
Workstation software.
module
subscribe sync
event exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
2. If not work, re-install
Workstation software.
module
unsubscribe
sync event
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
2. If not work, re-install
Workstation software.
module
unregister
servant
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
2. If not work, re-install
Workstation software.
controller
exception
Zoom Roam init exception.
Please restart the system. If
this issue persists, please call
for service.
2. If not work, re-install
Workstation software.
exception
exception. Display may be
abnormal. Please restart the
system. If this issue persists,
please call for service.
2. If not work, re-install
Workstation software.