Troubleshooting
6-35
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0310509
Warning
Stop
application
exception
Close Zoom Roam exception.
Please restart the system. If
this issue persists, please call
for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0310510
Warning
Start
zoomroam
exception
Open Zoom Roam exception.
Please restart the system. If
this issue persists, please call
for service.
0310511
Warning
Stop zoomroam
exception
Close Zoom Roam exception.
Please restart the system. If
this issue persists, please call
for service.
0310512
Warning
Get string failed
Loading message fail. Please
restart the system. If this issue
persists, please call for service.
0310513
Error
ZoomRoam
module get
OSGI service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0310514
Warning
Get resource
bundle failed
Loading message fail. Please
restart the system. If this issue
persists, please call for service.
0311501
Warning
Read
configuration
file failed
Configuration loading failure.
Please restart the system.
Re install software or import
configure file
0311502
Warning
Write
configuration
file failed
Configuration save failure.
Some configuration may be
lost. Please restart the system.
Re install software or import
configure file
0311503
Error
Administration
module register
servant
exception
Workstation communication
exception. Please restart the
system.
Re install software
0311504
Error
CORBA
communication
failed
Workstation communication
exception. Please restart the
system.
Re install software

Table of Contents