Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-36
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0311505
Error
Administration
module
subscribe sync
event exception
Workstation communication
exception. Please restart the
system.
Re install software
0311507
Error
Thread
operation failed
Execution exception. Please
restart the system.
Re install software
0311508
Error
Administration
module register
OSGI service
exception
Workstation communication
exception. Please restart the
system.
Re install software
0311509
Error
OSGI
communication
failed
Workstation communication
exception. Please restart the
system.
Re install software
0311510
Error
Get authority
failed
Load account data failure.
Please restart the system.
Re install software
0311511
Warning
Get filter status
failed
Can’t get the filter option's
status. Please restart the
system.
Re install software
0311512
Warning
Get mainframe
status failed
Can’t get the mainframe
connection status. Please
restart the system.
Re install software
0311514
Error
Register CORBA
notify failed
Workstation communication
exception. Please restart the
system.
Re install software
0311519
Warning
File operation
exception
File operation exception.
Please retry. If this issue
persists, please restart the
system.
Re install software
0311520
Warning
Execute
command
failed
System execution exception.
Please try again. If this issue
persists, please call for service.
Re install software