Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-38
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
restart the system.
0311535
Warning
Boot up
terminal failed
Open terminal failure. Please
restart the system.
Re install software
0311536
Warning
Get un-finalized
image failed
Search unfinalized image
failure. Please restart the
system.
Re install software
0311537
Warning
Perform
calibration
failed
Calibration failure. Please retry.
If this issue persists, please call
for service.
Re install software
0311539
Warning
Calculated
Dose or DAP
coefficient
failed
Dose or DAP coefficient
calculation exception. Please
retry. If this issue persists,
please restart the system.
0311540
Warning
Compress
folder failed
Compress file failure. Please
retry. If this issue persists,
please restart the system.
0311541
Warning
Get string failed
Loading message fail. Please
restart the system. If this issue
persists, please call for service.
0311542
Warning
Get resource
bundle failed
Loading message fail. Please
restart the system. If this issue
persists, please call for service.
0311543
Error
Administration
module get
OSGI service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0311544
Error
Administration
module
unregister OSGI
service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0311545
Error
Administration
module
unregister
servant
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.