Troubleshooting
6-39
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0311546
Error
Administration
module
unsubscribe
sync event
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0309500
Warning
Setup.xml does
not exist
Setup configuration loading
failure. Please restart the
system. If this issue persists,
please call for service.
check if configuration file
crashed
0309501
Warning
Setup
Default.xml
does not exist
Default Setup configuration
loading failure. Please restart
the system. If this issue
persists, please call for service.
check if default
configuration file crashed
0309502
Warning
ConnectivitySet
up.xml is not
exist
Connectivity configuration
loading failure. Please restart
the system. If this issue
persists, please call for service.
check if Dicom
configuration file crashed
0309503
Warning
ConnectivitySet
upDefault.xml is
not exist
Default connectivity
configuration loading failure.
Please restart the system. If
this issue persists, please call
for service.
check if default Dicom
configuration file crashed
0309504
Warning
Administration.
xml is not exist
Administration configuration
loading failure. Please restart
the system. If this issue
persists, please call for service.
check if admin
configuration file crashed
0309505
Warning
ExposureContro
l.xml is not exist
Exposure control configuration
file loading failure. Please
restart the system. If this issue
persists, please call for service.
check if exposure
configuration file crashed
0309506
Warning
SwBuild.xml is
not exist
Software version access
failure. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0309507
Warning
SerialNumber.x
ml is not exist
Serial number access failure.
Please restart the system. If
this issue persists, please call
for service.
1. Restart the system.
2. If not work, re-install
Workstation software.