Troubleshooting
6-53
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0405503
Warning
SCU performs
MPPS operation
failed
Report MPPS failure. Please
make sure the network and
MPPS server configuration are
correct, and the server is
available. And try again. If this
issue persists, please call for
service.
Check DICM configuration.
0405504
Warning
SCU performs
MWL operation
failed
Worklist query failure. Please
make sure the network and
worklist server configuration
are correct, and the server is
available. And try again. If this
issue persists, please call for
service.
1. Check worklist server
config tab, press verify
button, confirm connection
established
2. Confirm AE Title, IP,
hospital server settings
3. Confirm correct query
date range
4. Reinstall workstaion
software
0405505
Warning
SCU performs
query operation
failed
DICOM query failure. Please
make sure the network and
Q&R server configuration are
correct, and the server is
available. And try again. If this
issue persists, please call for
service.
Check DICM configuration.
0405506
Warning
SCU performs
retrieve
operation failed
DICOM retrieve failure. Please
make sure the network and
Q&R server configuration are
correct, and the server is
available. And try again. If this
issue persists, please call for
service.
Check DICM configuration.
0405507
Warning
SCU performs
storage
commitment
failed
DICOM storage commitment
failure. Please make sure the
network and storage
commitment server
configuration are correct, and
the server is available. And try
again. If this issue persists,
please call for service.
Check DICM configuration.