Troubleshooting
6-55
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0405707
Warning
SCU module
register servant
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service
make sure if omniNames
service failed, any module
crashed
0405708
Error
SCU inits a
semaphore
failed
Semaphore exception. Please
restart the system. If this issue
persists, please call for service.
system unknown error
0405709
Error
SCU module
unregister
servant
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service
0407501
Error
MM module
resolve object
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service
Make sure if omniNames
service failed, any module
initialization failed.
0407502
Error
MM module
Invoke corba
service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service
Make sure if omniNames
service failed, check
omniOrb log file.
0407503
Warning
Cancel thread
failed
Cancel execution thread
exception. Please restart the
system. If this issue persists,
please call for service.
System error: cant cancel a
thread.
0407504
Error
MM module
register servant
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service
Make sure if omniNames
service failed, check
omniOrb log file.
0407505
Error
Init semaphore
failed
Semaphore exception. Please
restart the system. If this issue
persists, please call for service.
Check if the semaphore is
valid, or check if the
maximum allowable value
for the semaphore will be
exceeded.
0407506
Error
Set locale failed
System start failure. Set locale
failed. Please restart the
system. If this issue persists,
please call for service.
System error. Call to
setlocal() function failed.
0407507
Error
Allocate
memory failed
System run out of memory.
Please restart the system. If
this issue persists, please call
for service.
Check memory using state.