Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-56
ID
Warning
Description
(Displayed in User Interface)
devices failed
Enumerate devices failure.
Please restart the system. If
this issue persists, please call
for service.
enumerate devices.
script failed
Create CD/DVD file exception.
Please retry or insert another
disk. If this issue persists,
please call for service.
script is the exact version
that is shipped with
product.
script failed
Copy to USB disk exception.
Please check the USB disk or
replace with another disk. If
this issue persists, please call
for service.
script is the exact version
that is shipped with
product.
application
failed
System module init failure.
Please restart the system. If
this issue persists, please call
for service.
Make sure there is enough
memory and the udev
library is correctly installed.
failed
MediaManagement report
ready failed. Please restart the
system. If this issue persists,
please call for service.
Application can’t be booted,
contact service.
failed
failed. Please try again. If this
issue persists, please call for
service.
System error. Call to system
function failed.
temperature
exception
exception. Please restart the
local printer and retry. If this
issue persists, please call for
service.
Check the temperature for
the printer.
feed error
Paper feed error. Please take
the paper out of printer and
put it in again. If this issue
persists, please call for service.
Get the paper out of printer
and put it in.
initialized
Printer is not ready. Please
make sure the printer is well-
connected and already turned
on. And then try again. If this
issue persists, please call for
service.
Make sure the printer has
been initialized.