Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-60
ID
Warning
Description
(Displayed in User Interface)
error
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service
1. Software internal error
2. Corba server error.
setup
configuration
file failed
Setup configuration loading
exception. Some setup
configuration may be missed.
Please restart the system. If
this issue persists, please call
for service.
2. Configure file format
error
administration
configuration
file failed
Administration configuration
loading exception. Please
restart the system. If this issue
persists, please call for service.
2. Configure file format
error
version
configuration
file failed
failure. Please restart the
system. If this issue persists,
please call for service.
2. Configure file format
error
the message
Unexpected message. Pleasre
restart the system. If this issue
persists, please call for the
service.
workstation
softawre
incompatible
incompatible. Mainframe is
stopped. Please make sure
connecting the right
mainframe. If the mainframe is
correct, please call service to
upgrade mainframe and/or
workstation software.
software version are not
compatible.
1. Check MCB input power
voltage.
2. Reinstall computer
software.
3. Reinstall MCB software.
connectivity
setup
configuration
file failed
Connectivity configuration
loading failure. Some
configuration may be missed.
Please restart the system. If
this issue persists, please call
for service.
module get
OSGI service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.