Troubleshooting
6-61
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0412514
Error
SystemControl
module
unregister OSGI
service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0412800
Warning
Serial numbers
of MCB and
WKS are not
compatible
Mainframe has different serial
number from Workstation.
Mixing use of workstation with
other system's mainframe is
not supported. Please make
sure workstation and
mainframe is match.
0412801
Warning
Static image
storage space
is full
Image storage capacity limit of
100,000 images reached.
Can’t start Digital Spot
exposure. Please delete
useless or out-of-date images
to free disk space.
0412804
Warning
Product models
of MF and WKS
are different
Workstation product model is
different from mainframe.
Mixing use of workstation with
different product model
mainframe is not supported.
Please make sure workstation
and mainframe is match.
0412805
Error
Workstation
product model
not configured
Product is not configured.
Please call service to configure
product model with product
license.
0412806
Warning
Workstation
product model
configured by
MFG
System does not have product
license. System can’t work with
full functions. Please call
service to configure product
with product license.
0412807
Warning
Get string failed
Loading message fail. Please
restart the system. If this issue
persists, please call for service.
0412808
Warning
Get resource
bundle failed
Loading message fail. Please
restart the system. If this issue
persists, please call for service.