Troubleshooting
6-67
ID
Warning
Description
(Displayed in User Interface)
UPS
Please keep AC plugged in to
avoid data loss. If this issue
persists. Please call for service.
UPS and WKS is abnormal.
1. PC USB for UPS may be
abnormal
2.connection between PC
and UPS may be abnormal
3. UPS may be abnormal.
4. Reinstall WKS software
be replaced for
aging of battery
UPS battery aged. Please call
service to replace the battery.
overloads
UPS battery overloaded. Please
restart the system to avoid
data loss. If this issue persists,
please call for service.
1, PC power supply may be
abnormal
failed
Application modules boot up
failure. Please restart the
system. If this issue persists,
please call for service.
up failed
failure. Please restart the
system. If this issue persists,
please call for service.
communication
failed
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
resolve object
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
Invoke corba
service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.