Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-66
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
4. PDU may be abnormal.
0408007
Warning
Secondary
voltage
underload
Secondary voltage underload.
It may result in exposure
exception. Please restart the
system. If this issue persists,
please call for service.
1. Main transformer
primary input power supply
may be abnormal.
2. 28V output of main
transformer may be
abnormal.
3. Adjust R52 follow SM
chapter 4.9
4. PDU may be abnormal.
0408008
Warning
PDU not detect
IPC 12V
Power monitoring exception. If
this issue persists, please call
for service.
1. 12V output of WKS may
be abnormal.
2. 12 V input for PDU may
be abnormal.
0408009
Error
Open serial port
failed
Workstation and peripheral
interface communicaiton
exception. Please restart the
system. If this issue persists,
please call for service.
Serial port of WKS abnormal
1. WKS serial port for UPS
may be abnormal.
2.Reinstall WKS software
0408010
Warning
Write serial port
failed
Workstation and peripheral
interface communicaiton
exception. Please restart the
system. If this issue persists,
please call for service.
Serial port of WKS abnormal
1. WKS serial port for UPS
may be abnormal.
2.Reinstall WKS software
0408011
Warning
Read serial port
failed
Workstation and peripheral
interface communicaiton
exception. Please restart the
system. If this issue persists,
please call for service.
Serial port of WKS abnormal
1. WKS serial port for UPS
may be abnormal.
2.Reinstall WKS software
0408012
Error
Set serial port
failed
Workstation and peripheral
interface communicaiton
exception. Please restart the
system. If this issue persists,
please call for service.
Serial port of WKS abnormal
1. WKS serial port for UPS
may be abnormal.
2.Reinstall WKS software
0408015
Warning
No battery in
UPS
No battery in UPS. Please keep
AC plugged in to avoid data
loss. If this issue persists.
Please call for service.

Table of Contents