Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
7-28
3. Put the Workstation brake pedal to position 1. Remove the screws of brake pedal and remove the
brake pedal.
4. Loose the nut and remove the central control wheel.
5. Then replace the wheel with a new one.
Pay attention to keep the direction and gear of
the new wheel is same as the original one.
6. Install the brake pedal as original position.
7. Place the brake pedal in shift 1/2/3 and verify the central
control wheels are functional in each shift. Place the brake
pedal in each shift and verify the related function, refer to
Section 3.6.1.