Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
7-36
5. Refer to calibration matrix in the appendix to perform calibration and checks.
After calibration and check, shut down system and reinstall the PCB Box and mainframe back covers.
Mother board (P/N: 5085601)
1. Remove the rear cover of the C-Arm console.
2. Release the screws of e-Box cover and remove it.
3. Remove the collimator control board, mainframe control
board and Filament Drive Board.
4. Release the screws and remove the whole e-Box.
5. Unplug all cables connected to the motherboard.
6. Loose the screws and remove the mother board.
7. Remove 3 PWA insert error proofing bolts from defect
mother board and assemble them on new board.
8. Assemble the new mother board and connect the cables
according to the label of the cable and the mark on the
board.