Replacement
7-35
Warning 0117800 may occur when Brivo OEC 715/785/865 system boot up with a new
MCB installed. Do not shut down the system by pressing the confirm button. Instead, hold
down Press keys “Ctrl +Shift+Alt+
→
” to invoke the service tool first in Administration
interface. Then, refer to the chapter 4 and follow the instructions to finish the calibrate
process.
For MCB with Firmware 1.9.0-F or lower version, when replacing new version MCB, WKS
APP SW should be also upgraded to WKS 1.0.
Filament Drive Board (P/N: 5085604)
1. Remove the rear cover of the C-Arm console.
2. Release the screws of PCB Box cover, remove Filament Diver Board and insert new one.
3. Configure the Jumper: J5 for old Monoblock without blue mark or J6 for new Monoblock
with Blue Mark:
The jumper is installed on J6 for new Monoblock as default.
4. Confirm the switch J7 is set to small position. If not, configure to small focal spot.