Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
7-72
10. Remove the screws on the top end of external column, and remove the side cover. Remove the cap
from the inner column. Fasten a M4 screw into the screw hole on actuator upper shaft and draw the
shaft out.
Upper shaft structure:
Fasten M4 screw here and
pull out upper shaft
Upper Shaft
Lower Shaft