Brivo OEC 715/785/865 Mobile C-Arm X-ray Product Service Manual
8-12
C-Arm Controls Check:
Refer to section 3.6.2 for C-Arm Controls Check.
Perform the following C-Arm control checks:
Some of the following procedures produce X-rays. Take appropriate precautions.
X-ray Key/X-ray Switch Functional Test:
Perform the following test to verify that the system’s X-ray and key switches are operating satisfactorily.
This procedure produces X-rays. Take appropriate precautions.
1. Place the X-ray Keyswitch in STANDBY position ( ). Press Handswitch and verify that no X-rays are
generated. Press Vertical lift button and verify that vertical lift column is not operational.
2. Place X-ray Keyswitch in XRAYOFF position ( ). Press Handswitch and verify that no X-rays are
generated. Press vertical lift button and verify the vertical lift column is operational.
3. Place X-ray Keyswitch in ON position ( ). Press Handswitch/Footswitch and vertical lift button to
verify X-ray and vertical lift column are enable.
Emergency Switch:
Perform the following test to verify the emergency switch.
1. Press the emergency switch during raising C-
arm.
2.
Verify that vertical column operation and
exposure are disabled.
3. Release the emergency switch. Verify that the
vertical column operation and exposure are
available.