EasyManua.ls Logo

GE Brivo OEC 715 - Page 68

GE Brivo OEC 715
551 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Brivo OEC 715/785/865/865 Mobile C-Arm X-Ray Product Service Manual
2-14
MF_PWR
mainframe power detection
signal
PJ8-8 digital 0/+12V (>6V=High)
PDU_CONNECT_2
connection status signal
from PDU to SIB
PJ8-9 digital 0/+12V (>6V=High)
PDU_CONNECT_1
connection status signal
from PDU to SIB
PJ8-10 digital 0/+12V (>6V=High)
RSRV_FROM_PDU_1
Reserved signal from PDU
PJ8-11
digital
0/+12V (>6V=High)
RSRV_FROM_PDU_2
Reserved signal from PDU
PJ8-12
digital
0/+12V (>6V=High)
RSRV_FROM_PDU_3
Reserved signal from PDU
PJ8-13
digital
0/+12V (>6V=High)
GND
GND
PJ8-14
GND
0V
RSRV_FROM_SIB_1
Reserved signal from SIB
PJ8-15
digital
0/+12V (>6V=High)
RSRV_FROM_SIB_2
Reserved signal from SIB
PJ8-16
digital
0/+12V (>6V=High)
RSRV_FROM_SIB_3
Reserved signal from SIB
PJ8-17
digital
0/+12V (>6V=High)
GND
GND
PJ8-18
GND
0V
BEEP_INVALID_KEY_C
MD
Invalid key beep command PJ8-19 digital 0/+12V (>6V=High)
INTLC_BOARD
interlock board connection
status signal
PJ8-20 digital 0/+12V (>6V=High)
ACK_PDU2SIB
acknowledge signal
between PDU and SIB
PJ8-21 digital 0/+12V (>6V=High)
SYS_PWR_OFF_CMD
system power off
command
PJ8-22 digital 0/+12V (>6V=High)
SYS_PWR_OFF_DENY
system power off deny
command
PJ8-23 digital 0/+12V (>6V=High)
MF_PWR_ON_CMD
mainframe power on
command
PJ8-24 digital 0/+12V (>6V=High)
RDY
WKS boot up ready
PJ8-25
digital
0/+12V (>6V=High)
BEEP_ERR_CMD
error beep command
PJ8-26
digital
0/+12V (>6V=High)
UPS interface
UPS switch on/off control signal interface is shown as follows:
Name
Description
Pin
Attribution
Scale
Comment
UPS_OUTPUT
UPS output status signal
PJ14-1
Analog
0/10-14VDC
GND
GND
PJ14-2
0
0
ON_BATT
UPS battery/line status signal
PJ14-3
Analog
0/10-14VDC
REMOTE
UPS remote signal
PJ14-4
Digital
3.3V/5V TTL
UPS switch off interface in emergency is shown as follows:
Name
Description
Pin
Attribution
Scale
Comment
EMERGENCY_
BUTTON_1
UPS EMERGENCY SWICH
OFF BUTTON
PJ15-1 Digital 3.3V/5V TTL
EMERGENCY_
BUTTON_2
PJ15-2 Digital 3.3V/5V TTL
Test Point:
Name
Description
Description
TP1
+12V
+12V power on board
TP2
+24V
+24V power on board
TP3
+5V
+5V power on board
TP4
+3.3V
+3.3V power on board
TP5
+10V
+10V power on board

Table of Contents