8-76 Section 8-5 - Back End Parts Replacement
8.) From the rear of the BEP2, remove the four screws that fasten the power supply in position on the
BEP, as shown in Figure 8-59.
Note: For the next steps, it is recommended to carefully lie the BEP2 on its side, to provide easier
access.
9.) Cut any cable ties holding the power supply cable in place, as shown in Figure 8-60 below.
Figure 8-59 BEP2 Rear View
Figure 8-60 BEP2 Power Supply and Cable Ties
Power Supply
Cut Cable Ties Here
Cut Cable Ties Here