GE MEDICAL SYSTEMS 
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
Chapter 8 - Replacement Procedures  8-77
10.)Gently slide the power supply a short distance backwards (away from the BEP casing), while tilting 
it downwards and pulling it to the side.
NOTE: For clarification, the numbers appearing in square brackets in the following steps correspond to the 
numbers shown in the illustration of the BEP2 Power Supply and cable connectors in Figure 8-61 below.
11.)Carefully remove the ECG Module from the front of the BEP, sliding it out and disconnecting the 
power cable [1] from the rear of the ECG Module. (Refer to the BEP2 - ECG Module Removal 
Procedure on page 8 - 70).
12.)Disconnect the power cables from the media drives (MOD [2] and CD-RW [3]).
13.)Remove the screw that fastens the Ground cable and PC-VIC in position, as shown in Figure 8-62 
below.
14.)Disconnect the power cable from the Hard Disk (labelled [4] in Figure 8-61); refer also to Figure 8-62.
15.)Disconnect the data cable from the PC-VIC (refer to Figure 8-62, above).
16.)Slide the PC-VIC & hard disk out sufficiently to be able to cut any cable ties and disconnect the 
power cable (labelled [5] in Figure 8-61) from the PC-VIC; refer also to Figure 8-62, above. 
Figure 8-61 BEP2 Power Supply with Cable Loom and Connectors
Figure 8-62 Disconnecting Cables
1
3
4
9
6
7
8
Not used
Cable Connections
1 ECG
2 MOD
3 CD-RW
4 Hard Disk
5 PC-VIC
6 Floppy Disk
7 BEP rear
8 Motherboard (beside ECG)
9 Motherboard (beside CPU)
2
5
Ground Cable Screw
Power Cable to PC-VIC
Data Cable to PC-VIC
Power Cable to Floppy Disk
Power Cable to Hard Disk
PC-VIC