7.1 Saving the Test Conditions
101
7
Screen for With-
stand-voltage
test
Switch for test-voltage value, test frequency and initial ramp-up value
→ key
Switch for upper-limit test value and lower-limit test value → key
Switch for test time, ramp-up time and ramp-down time → ON/OFF key
Switch for test-voltage value and contact check upper (lower)-limit value
→ LOCK key (LOCAL key)
Screen for Insu-
lation-resis-
tance test
Switch for upper-limit test value and Lower-limit test value → key
Switch for test time and delay time → ON/OFF key
Switch for test-voltage value and contact check upper (lower)-limit value
→ LOCK key (LOCAL key)
Test type “AC” lights up during withstanding testing.
Test-voltage value Saved test voltage.
Upper-limit value icon Lights up when indicating the upper-limit test value.
Lower-limit value icon Lights up when indicating the lower-limit test value.
Upper-limit test value
Lower-limit test value
Lit up upper-limit icon indicates saved upper-limit test
value, lit up lower-limit icon indicates saved lower-limit
test value.
Test time
Ramp-up time
Ramp-down time
Indicates saved test time, ramp-up time and ramp-
down time.
File No. Indicates the File No. of saved test conditions.
File No.
Test type
Lower-limit value icon
Test-voltage value
Upper-limit test value
Lower-limit test value
Test time
Ramp-up time
Ramp-down time
Test-voltage value Saved test voltage.
Upper-limit value icon Lights up when indicating the upper-limit test value.
Lower-limit value icon Lights up when indicating the lower-limit test value.
Upper-limit test value
Lower-limit test value
Lit up upper-limit icon indicates saved upper-limit test
value, lit up lower-limit icon indicates saved lower-limit
test value.
Test time
Delay time
Indicates saved test time and delay time.
File No. Indicates the File No. of saved test conditions.
File No.
Test-voltage value
Lower-limit value icon
Upper-limit test value
Lower-limit test value
Test time
Delay time