9.4 Message Reference
144
Queries device ID.
Syntax
(Query)
IDN?
Queries manufacturer's name, model name, and software version.
(Response)
<Manufacturer's name>,<Model name>,0,<Software version>
Example
(Query)
IDN?
(Response)
HIOKI,3174,0,V1.00
The Device ID is HIOKI, 3174, Model code 0 (standard model), software ver-
sion 1.00.
Note
A header is not added to the response message.
Initialize Device
Syntax
(Command)
RST
This resets all settings, apart from saved data, to the default factory
settings.
After initialization, the screen returns to the initial screen.
Note Send the :SYStem:RESet command if initialization of saved data is
required.
The settings of the interface command "START", communication conditions
(the PC interface, GP-IB address) and the inter-lock function cannot be reset.
The memory of the set values cannot be reset.
For items to be initialized, refer to "11.4 System Reset" (p. 198)
Execute Self-Test and Query the Result
Syntax
(Query)
TST?
Perform instrument self-test and return the result as numerical value 0
to 3.
(Response)
<0 to 3 (NR1)>
0....... No Errors
1....... RAM Error
2....... EEPROM Error
3....... RAM Error and EEPROM Error
Example
(Query)
TST?
(Response)
1
A RAM error occurred.
Error
The execution of this command in a state other than the READY state causes an
execution error.
Note
A header is not added to the response message.