9.4 Message Reference
166
Sets and Queries the test-voltage value for insulation-resistance tests
Syntax
(Command)
:CONFigure:INSulation:VOLTage <500,1000 (NRf)>
Sets the test-voltage value (unit: V) for insulation-resistance tests, in
the READY state.
(Query)
:CONFigure:INSulation:VOLTage?
Queries the test-voltage value (unit: V) for insulation-resistance tests.
(Response)
<500,1000 (NR1)>
Example
(Command)
:CONFigure:INSulation:VOLTage 500
Sets the test-voltage value for the insulation-resistance tests to 500 V.
(Query)
:CONFigure:INSulation:VOLTage?
(Response)
500
The test-voltage value is 500 V.
Error
• The execution of this command in a state other than the READY state causes
an execution error.
• Execution error appears when value other than 500 or 1000 is sent.
Sets and Queries the upper-limit test value for insulation-resistance tests
Syntax
(Command)
:CONFigure:INSulation:RUPPer <0.20 to 2000 (NRf)>
Sets the upper-limit test value (unit: MΩ) for insulation-resistance
tests, in the READY state.
(Query)
:CONFigure:INSulation:RUPPer?
Queries the upper-limit test value (unit: MΩ) for insulation-resistance
tests.
(Response)
<0.20 to 2000 (NR1/NR2)>
Example
(Command)
:CONFigure:INSulation:RUPPer 2000
Sets the upper-limit test value for the insulation-resistance tests to 2000 MΩ.
(Query)
:CONFigure:INSulation:RUPPer?
(Response)
2000
The upper-limit test value is 2000 MΩ.
Error
• The execution of this command in a state other than the READY state or "test
setting state" causes an execution error.
• When the upper limit of the insulation-resistance test is set to ON, an execution
error will occur when “upper limit test value < lower limit test value” is set.
In addition, when the upper limit of the insulation-resistance test is set to ON,
and the auto range of the insulation resistance test is set to fixed range in the
optional function setting, an execution error will occur when a value exceeding
the resistance measurement range is set.
When the test-voltage is 1000 V, values below 1.00 MΩ are outside the resis-
tance measurement range.
Refer to "4.1.2 Setting the Upper (Lower) Limit Value" (p. 51) for resistance
measurement ranges.