EasyManua.ls Logo

Hioki 3174 - Page 175

Hioki 3174
224 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9.4 Message Reference
169
99
Enables/disables and Queries the test time for insulation-resistance tests
Syntax
(Command)
:INSulation:TIMer <ON/ OFF>
Enables and disables the test time for insulation-resistance tests, in
the READY state.
(Query)
:INSulation:TIMer?
Queries the test time enablement for insulation-resistance tests.
(Response)
<ON/ OFF>
Example
(Command)
:INSulation:TIMer ON
Enables the test time for an insulation-resistance test.
(Query)
:INSulation:TIMer?
(Response)
OFF
The test time is set to OFF.
Error
The execution of this command in a state other than the READY state or "test set-
ting state" causes an execution error.
Enables/disables and Queries the delay time for insulation-resistance tests
Syntax
(Command)
:INSulation:DELay <ON/ OFF>
Sets the delay time ON/OFF for insulation-resistance tests in the
READY state.
(Query)
:INSulation:DELay?
Queries the delay time ON/OFF for insulation-resistance tests.
(Response)
<ON/ OFF>
Example
(Command)
:INSulation:DELay ON
Sets the delay time for insulation-resistance tests to ON.
(Query)
:INSulation:DELay?
(Response)
OFF
The delay time is set to OFF.
Error
The execution of this command in a state other than the READY state or "test set-
ting state" causes an execution error.