9.4 Message Reference
168
Sets and Queries the delay time for insulation-resistance tests
Syntax
(Command)
:CONFigure:INSulation:DELay <0.1 to 99.9 (NRf)>
Sets the delay time (unit: s) for insulation-resistance tests in the
READY state.
(Query)
:CONFigure:INSulation:DELay?
Queries the delay time (unit: s) for insulation-resistance tests.
(Response)
<0.1 to 99.9 (NR2)>
Example
(Command)
:CONFigure:INSulation:DELay 3.0
Sets the delay time for insulation-resistance tests at 3.0 s.
(Query)
:CONFigure:INSulation:DELay?
(Response)
3.0
The delay time is set to 3.0 s.
Error
The execution of this command in a state other than the READY state or "test set-
ting state" causes an execution error.
Enables/disables and Queries the upper-limit test value for insulation-resistance tests
Syntax
(Command)
:INSulation:RUPPer <ON/ OFF>
Enables and disables the upper-limit test value for insulation-resis-
tance tests, in the READY state.
(Query)
:INSulation:RUPPer?
Queries the upper-limit test value enablement for insulation-resis-
tance tests.
(Response)
<ON/ OFF>
Example
(Command)
:INSulation:RUPPer ON
Enables the upper-limit test value for an insulation-resistance test.
(Query)
:INSulation:RUPPer?
(Response)
OFF
The upper-limit test value is set to OFF.
Error
The execution of this command in a state other than the READY state or "test
setting state" causes an execution error.
When the upper limit and lower limit values of the insulation-resistance test are
set as “upper limit test value < lower limit test value”, an execution error will
occur when the upper limit of the insulation-resistance test is set to ON.
In addition, when the auto range of the insulation-resistance test is set to fixed
range in the optional function setting, an execution error will occur when the
upper limit test value exceeds the resistance measurement range when the
upper test limit is set to ON.
When the test-voltage is 1000 V, values below 1.00 MΩ are outside the resis-
tance measurement range.
Refer to "4.1.2 Setting the Upper (Lower) Limit Value" (p. 51) for resistance
measurement ranges.