QDR
4500 Technical Manual
6-22
Could not find file extension in
Software Error: This message should be reported to Hologic Technical Support. Include as
much information as possible about what you were doing at the time.
Couldn't write the new record
Hardware or
Software Error: The system was unable to write a reference curve that was entered or edited.
Action: Check that the hard drive is not full. That is, check the message "Room for
__ scans" in the status window and ensure that it does not say zero. If it is
full, then archive and delete one or more scans and try again.
If the hard drive is not full, then you have either a hard drive failure or a
corrupt reference curve database.
db_File error __ .....
or
db_VISTA error __ .....
Software Error: Any message that begins this way should be reported to Hologic Technical
Support. Include as much information as possible about what you were
doing at the time.
Device already in use by foreground task [E_LOCKOUT_I]
LOCUS: MAPI The MAPI driver was already opened by another application when the
driver attempted to go ON_LINE (i.e., the driver requires exclusive access
to the MAPI driver).
Action: Report this error to Hologic Technical Support.
LOCUS: STATE An application attempted to load or execute a state machine when one was
already running, or an application attempted to execute a state machine
while one or more devices were still executing a command, or an attempt to
abort the currently executing state machine failed.
Action: Report this error to Hologic Technical Support.
LOCUS: Motor_xx An application attempted to modify stepper motor parameters while the
motor was moving (acceleration distance/type, first/final rates, position
average, etc.), or an application issued a MOVE_ABS while a MOVE_REL
was in progress (or vice versa).