QDR
4500 Technical Manual
6-26
manually push the table and C-arm away from each other and then follow
the recovery procedure under E_COLLISION_IMMINENT_I.
End of file
Action: This could occur at the start of a lateral scan if the currently selected scan is
not an analyzed companion AP scan.
Error closing optical file ________________________
or
Error flushing optical file _______________________
or
Error writing optical record ______________________
Hardware error: There is a problem writing an optical file. Either the medium is full, or
there is a hardware malfunction.
Action: Check for disk full. Go to DOS, type DIR D: and check for free space.
If the disk is not full, call Hologic Technical Support.
Error copying file _________
Hardware Error: There was an error copying the named file.
Action: Verify that there was space on the target drive. Check the "Room for .. "
message in the status window. Run appropriate diagnostics.
Error creating file ______________ _________________
Hardware or
Software Error:
There is a problem creating a new file (which may be on your hard drive, your diskette,
or your optical). Either the medium is full, or there is a hardware error.
Action: First determine which disk drive is involved. The filename should start with
a drive letter, followed by a colon. Drive A: is diskette, C: is hard drive,
and D: is the optical.
Check for disk full. For diskette, simply try a different diskette and see if
the problem goes away. For hard drive, check the "Room for ___" message
in the status window. For optical, type DIR D:.
If the disk is not full, run appropriate diagnostics.