Section 6 - Fault Isolation
6-27
Error initializing graphics module.
Action: This can be caused by not enough available memory or an incorrect graphics
adapter. Verify that
only
Hologic software is installed on the system and
that the graphics adapter has not been modified.
Error in region of interest structure
Software Error: This message should be reported to Hologic Technical Support. Include as
much information as possible about what you were doing at the time.
Error opening the flattening input file
Action: Disk error or missing software. Check operation of hard disk, and check
consistency of system software with CHEKPART.
Error reading flattening configuration record
Action: Disk error or missing software. Check operation of hard disk, and check
consistency of system software with CHEKPART.
Error reading flattening factor record
Action: Most likely cause is no flattening done for given scan mode. Re-run
BIGFLAT.
Error reading flattening records
Action: Disk error or missing software. Check operation of hard disk, and check
consistency of system software with CHEKPART.
Error reading Optical drive: _______________
Error reading file _______ : __________
Hardware Error: There was a disk problem while reading a file.
Action: Run appropriate diagnostics.
Error reading reference attenuation record
Action: Flattening may not have been done for this scan mode. Re-run BIGFLAT.