QDR
4500 Technical Manual
6-28
Failure at file ________ line ____
Software Error: This message should be reported to Hologic Technical Support. Include as
much information as possible about what you were doing at the time.
file extension doesn't start with 'P' in ...
Software Error: This message should be reported to Hologic Technical Support. Include as
much information as possible about what you were doing at the time.
flattening records are not initialized
Action: Disk error or missing software. Check operation of hard disk, and check
consistency of system software with CHEKPART.
Garbled Optical File
Software Error: The optical file that you are attempting to restore scans from seems to be
clobbered.
Action: Run diagnostics on the optical drive. If this error recurs, report it to Hologic
Technical Support.
GetDKernel Internal Software Error
or
GetKernel Internal Software Error
Software Error: These messages should be reported to Hologic Technical Support. Include
as much information as possible about what you were doing at the time.