Section 6 - Fault Isolation
6-29
Histogram Overflow in datahist
or
Histogram Smoothing Error
or
Illegal Context Record Type
or
Illegal high value in qgen
or
Inconsistent d0 limits in rsattencalc
or
Indeterminate or bad data for attenuation curves
or
Indeterminate Data For k Calculation
or
Insufficient Data To Determine k/delta0
Software Error: These messages should be reported to Hologic Technical Support. Include
as much information as possible about what you were doing at the time.
Internal Buffer Size Exceeded
Software Error: This message should be reported to Hologic Technical Support. Include as
much information as possible about what you were doing at the time.
Internal device request queue full [E_QUEUE_FULL_I]
Comment: The common buffer area for storing system sounds is full.
Action: Run CHEKPART to verify system software is correctly installed. Reduce
the size of one or more of the tunes or tune files you used to specify the
system sounds.