6-30
Internal driver error [E_INTERNAL_ERROR_I]
Comment: If returned in response to an ON_LINE command, indicates that the MAPI
device driver is not properly installed in the system.
Action: Verify that the line DEVICE=C:\MENU\MAPIDEV.SYS is in the system
CONFIG.SYS file, that the interrupt level (IRQ) and I/O base address are
correctly set and that the file C:\MENU|MAPIDEV.SYS is installed on the
system.
Comment: If returned in response to a XRAY_ON_PHASE, XRAY_SEQUENCE, or
PULSE_PER_SEG command to the C-arm device, indicates that the
currently selected X-ray mode is invalid.
Action: Report this to Hologic Technical Support.
Comment: If reported as the device status for the C-arm, it indicates an unrecognized
response from the C-arm micro.
Action: This may be due to a communications error between the AT and the C-arm.
Follow the procedure under "BAD CHECKSUM IN MESSAGE." to
diagnose I/O errors.
Comment: If reported as the abort code for the state machine, it indicates that the
internal driver event queue was full when the state machine attempted to
issue the system-wide abort code for state machine termination.
Action: Report this error to Hologic Technical Support.
Internal Error:.....
or
Internal Software Error: ......
Software Error: Any message that begins this way should be reported to Hologic Technical
Support. Include as much information as possible about what you were
doing at the time.
Internal stack error [E_STACK_ERROR_I]
Comment: Returned by the state machine driver to indicate that the local stack for a
task has overflowed or underflowed.
Action: Report this problem to Hologic.