Section 6 - Fault Isolation
6-31
Invalid ...
Software Error: In general, any message that begins with the word "Invalid" should be
reported to Hologic Technical Support. Include as much information as
possible about what you were doing at the time. See the following
"Invalid..." errors for more information.
Invalid address for message packet [E_INVALID_ADDR_I]
Comment: The address of a message packet was invalid. The MAPI driver keeps an
address table that maps each byte address to its respective communications
link (SSI, SCI or C/C). The selected address was not assigned to any link
and therefore was not recognized by the MAPI driver.
Action: Report this error to Hologic Technical Support.
Invalid flattening data [E_INVALID_FLAT_DATA_S]
Comment: The most likely cause is that the signal for one or more of the detectors is
higher than allowed.
Action: One or more detectors is bad. See instructions for running hardware
checking program.
Invalid identification file.
Action: QDR 4500 ID file C:\SQUID.DAT is not valid. Call Hologic Technical
Support.
Invalid machine type in ARRC.TXT.
Action: This is not an error if the machine is not yet calibrated.
Invalid mode for device driver [E_INVALID_MODE_I]
LOCUS: C-Arm Returned by the C-arm if an application attempts to change the X-ray mode
when X-rays are already on.
Action: Report this error to Hologic Technical Support.
LOCUS: STATE Returned as the device state if the C-arm microprocessor returned an
unrecognized response to the fast query.
Action: This may be due to a communications error between the AT and the C-arm.
Follow the procedure under "BAD CHECKSUM IN MESSAGE." to
diagnose I/O errors. Otherwise, report this error to Hologic Technical
Support.