QDR
4500 Technical Manual
6-32
LOCUS: Motor_xx For a stepper motor, the application attempted to move a motor in the
opposite direction in which it is currently moving (i.e., a motor's motion can
be extended in the same direction as the current motion but it cannot reverse
motion).
Action: Report this error to Hologic Technical Support.
LOCUS: STATE For the state machine driver, this error indicates that either the protocol
attempted to execute a command that is not supported; that an instruction
that requires a register reference did not provide one; or, more commonly,
an attempt was made to access the external data buffers and they were either
not available or were still being using by the application.
Action: If you are running any program other than standard Hologic software,
remove it and retry the scan (e.g., do not run a scan while connected to a
network or attempt to run a scan under Windows). Otherwise, report this
error to Hologic Technical Support.
Invalid parameter(s) [E_INVALID_PARAM_I]
Comment: The value of the parameter associated with a device command was out of
range.
Action: Report the application in error to Hologic Technical Support.
Invalid serial number in ARRC.TXT.
Action: This is not an error if the machine is not yet calibrated.
Invalid serial number in environ.bat file.
Action: Machine serial number is not valid. Call Hologic Technical Support.
Invalid values in bad detector map
Action: See instructions for running hardware checking program.
I/O Error ....
Hardware or
Software Error:
There was an error reading or writing a disk drive. The message should identify the
drive and say something about the nature of the error.