Section 6 - Fault Isolation
6-33
Action: If this is an "out of space" error, verify that there is room on the target drive.
Check the "Room for ..." message in the status window. Otherwise, run
appropriate diagnostics on the disk and controller; if the problem involves
diskette (Drive A:), try a different diskette.
Limit Exceeded: ......
Software Error: Any message that begins this way should be reported to Hologic Technical
Support. Include as much information as possible about what you were
doing at the time.
Local motion active [E_LOCAL_MOTION_I]
Comment: An application has attempted to enable or disable local motion when one or
more motors were still moving. This is also the abort code reported by the
state machine when a protocol is aborted to start the local motion state
machine.
Action: When moving the system using the operator control panel, wait until all
motion stops before starting an application that requires motor movement
(e.g., the scan program).
Machine serial number is not valid.
Action: QDR 4500 ID file C:\SQUID.DAT is not valid. Call Hologic Technical
Support.
Machine type in configuration files is inconsistent.
Action: QDR 4500 ID file C:\SQUID.DAT is not valid. Call Hologic Technical
Support.
Machine type is not valid.
Action: QDR 4500 ID file C:\SQUID.DAT is not valid. Call Hologic Technical
Support.
Memory allocation error.
Action: Verify that
only
Hologic software is installed on the system.