QDR
4500 Technical Manual
6-36
No motion detected while motor stepping [E_NO_MOTION_I]
Hardware Error: The QDR is equipped with motion detectors which detect actual motion in
the X, Y and C-arm rotation motors. This message occurs when one of
these motors was commanded to move and no motion was detected.
Action: Check for loose belt couplers, tighten set screws A/R.
Check for coupler hitting bearing block, move coupler A/R.
Check motion encoder, replace if defective.
Check Motor & Drive board, replace if defective.
Check stepper translator, replace if defective.
MOTOR$xx: Whenever a motor is moving, the driver checks its position every three
seconds and stops stepping if the incremental positions at these check
intervals do not match the motion parameters (final rate) to within
approximately 50%.
Action: Determine why the motor is not moving (check motor coupling, drive belt,
and encoder shaft coupling).
MOTOR$TZ: The device state reported whenever the TZ firmware reports a pedestal
timeout during a move command.
Action: See the discussion of the TZ drive's E_TOLERANCE_I error.
No data for attenuation curves
Software Error: This message should be reported to Hologic Technical Support. Include as
much information as possible about what you were doing at the time.
Not enough memory
Software Error: Ensure that
only
Hologic software is installed on the computer. If so, this
message should be reported to Hologic Technical Support. Include as much
information as possible about what you were doing at the time.
Old Whole Body Context Type
Software Error: This message should be reported to Hologic Technical Support. Include as
much information as possible about what you were doing at the time.
Operator terminated action [E_OPERATOR_I]
Action: None.