Section 6 - Fault Isolation
6-37
Out of memory
Software Error: Ensure that
only
Hologic software is installed on the computer. If so, this
message should be reported to Hologic Technical Support. Include as much
information as possible about what you were doing at the time.
Overflow calculating flattening or daily factors
Action: Most likely cause is saturation of one or more detectors during the flattening
procedure.
Parameter size is invalid [E_INVALID_SIZE_I]
Comment: The device state reported by the C-arm if the response to a fast query is not
a minimum length to encompass all necessary information.
Action: This may be due to a communications error between the AT and the C-arm.
Follow the procedure under "BAD CHECKSUM IN MESSAGE." to
diagnose I/O errors.
Parameter size is too large for driver [E_TOO_LONG_I]
Comment: For the state machine, indicates that a reference was made beyond the
machine limits (e.g., a register reference that exceeded the number of
registers; a jump to a location beyond the end of the state machine; etc.).
For all other devices, indicates that the message response from the remote
micro overflowed the internal buffers and was partially lost.
Action: For the state machine, report this error to Hologic Technical Support. For
other devices, this error may indicate a problem on the communications bus.
Follow the procedure under "BAD CHECKSUM IN MESSAGE." for
isolating I/O errors.
Patient File Record __ Too Small
or
Patient File Record __ Too Large
Hardware or
Software Error:
These messages indicate problems performing I/O operations (usually to the hard drive,
but possibly to diskette or optical).
Action: Ensure that there is adequate space on the hard drive (Check the "Room for
..." message in the status window. You should always leave room for 1 or 2
scans). There may be a disk problem; appropriate diagnostics should be
run. If you are unable to find a hardware problem, the message should be