QDR
4500 Technical Manual
6-38
reported to Hologic Technical Support. Include as much information as
possible about what you were doing at the time.
PGLINE: ...
Software Error: Any message that begins this way should be reported to Hologic Technical
Support. Include as much information as possible about what you were
doing at the time.
Remote device did not echo command [E_NAK_I]
Comment: Whenever the first byte in a message response does not match the command
character.
Action: This is probably a communications error between the AT and the
microprocessor. Follow the procedure under "BAD CHECKSUM IN
MESSAGE." to diagnose I/O errors.
ROI Limit Error
Software Error: This message should be reported to Hologic Technical Support. Include as
much information as possible about what you were doing at the time.
Scan File Record __ Too Small (__)
or
Scan File Record __ Too Large (__)
or
Scan has 0 points or lines
Hardware or
Software Error:
These messages indicate problems performing I/O operations (usually to the hard drive,
but possibly to diskette or optical).