Section 6 - Fault Isolation
6-39
Action: Ensure that there is adequate space on the hard drive (Check the "Room for
..." message in the status window. You should always leave room for 1 or 2
scans). There may be a disk problem; appropriate diagnostics should be
run. If you are unable to find a hardware problem, the message should be
reported to Hologic Technical Support. Include as much information as
possible about what you were doing at the time.
Software Error .......
Software Error: Any message that begins this way should be reported to Hologic Technical
Support. Include as much information as possible about what you were
doing at the time.
Starting phase Out of Range
Software Error: This message should be reported to Hologic Technical Support. Include as
much information as possible about what you were doing at the time.
Test failed - absolute max or min limit exceeded
Action: One or more detectors is bad. See instructions for running hardware
checking program.
Test failed - maximum SD exceeded
Action: Detectors are not seeing a uniform beam. Check all hardware having to do
with X-ray generation or detection.
Test failed - mean exceeds limits
Action: X-ray beam at edges is not consistent with beam at center. Check X-ray
beam alignment.