4500 Technical Manual
6-40
The analyze data file either did not exist or did not contain any
valid entries
Software Error: The software was not correctly installed (No ANALYZE.DAT file). This
might also be due to a hardware problem with the hard disk drive.
Action: Notify Hologic Technical Support.
There are no files available to restore (
Optical
restore)
User or
Software Error:
When restoring files from optical ("Optical" on main menu), you selected an archive
that contains no scans.
Action: Report this problem to Hologic Technical Support.
There are no records in the Normals Curve database.
Software Error: The reference curve database appears to be totally empty. This condition
should never occur in normal operations.
Action: The Hologic software may not have been installed correctly.
The X-ray controller is not generating A/C line interrupts
[E_NO_TIMER_INTERRUPT_I]
Hardware Error: The X-ray Controller assembly generates an interrupt every 1/120 of a
second (60HZ line) or every 1/100 of a second (50Hz line). The PC clock
generates 18 interrupts a second (regardless of line frequency). If two clock
ticks occur with NO Motor & Drive board interrupts between them, then this
message is produced.
Action: Report this error to Hologic Technical Support. Include as much
information as possible about what you were doing at the time.
Refer to the
No A/C Line Interrupts
header, page 6-17 in this section.