Section 6 - Fault Isolation
6-41
Truncated Patient Record ...
Hardware or
Software Error:
This messages indicate problems performing I/O operations (usually to the hard drive,
but possibly to diskette or optical).
Action: Ensure that there is adequate space on the hard drive (Check the "Room for
..." message in the status window. You should always leave room for 1 or 2
scans). There may be a disk problem; appropriate diagnostics should be
run. If you are unable to find a hardware problem, the message should be
reported to Hologic Technical Support. Include as much information as
possible about what you were doing at the time.
device:
Unable To Allocate Resources
Software Error: This message should be reported to Hologic Technical Support. Include as
much information as possible about what you were doing at the time.
Unable to Allocate Space for Environment
or
Unable to Restore Environment
or
Unable to Calculate Line-By-Line d0's
Software Error: These messages should be reported to Hologic Technical Support. Include
as much information as possible about what you were doing at the time.
Unable to determine power line frequency
Hardware Error: The C-Arm was unable to measure line frequency interval.
Action: Check function of the C-Arm.
Check stability of A/C line frequency.