4.2.6.1 Setup configuration: Sound
Access sequence: RUN>SETTINGS>SETUP>CONFIGURATION>SOUND
•
Select “2” to emit a sound when the touch screen is tapped;
•
Select “1” to emit a sound after the end of each single injection, every sample
during an automatic run, when “save” or “store” is tapped, or if the autosampler memory is
erased;
•
Select “0” to disable the sound emission.
Tap “SAVE” to store the modification, otherwise “CANCEL”.
Select “Exit” to exit from this screen and return to the “Setup Configuration”” screen
(see paragraph 4.2.6 “Setup: Configuration”):
4.2.6.2 Setup configuration: Light
Access sequence: RUN>SETTINGS>SETUP>CONFIGURATION>LIGHT
This is used to change the light setting
•
Select “On” to switch on the target illumination;
•
Select “Off” to switch off the target illumination;
•
Select: “Injection” to switch on the target illumination during injection
only; Tap “SAVE” to store the modification, or “CANCEL”.
Select “Exit” to exit from this screen and return to the “Setup Configuration” screen
(see paragraph 4.2.6 “Setup: Configuration“).
4.2.6.3 Setup configuration: Run
Access sequence: RUN>SETTINGS>SETUP>CONFIGURATION>RUN
Use this to change the following operational parameter:
•
Tray Behaviour: to set tray position during automatic run;
Tray
Closed: the tray remains closed during run
Open: the tray remains open during run
•
Power-on Restart: sets autosampler restart after a power failure. It should be
set to “ON” only if your GC/analyzer and your data acquisition system are also
able to restart automatically;