4.2.6.1 Setup configuration: Sound
Access sequence: RUN>SETTINGS>SETUP>CONFIGURATION>SOUND
Select 2” to emit a sound when the touch screen is tapped;
Select 1 to emit a sound after the end of each single injection, every sample
during an automatic run, when “save” or store” is tapped, or if the autosampler memory is
erased;
Select 0 to disable the sound emission.
Tap “SAVE to store the modification, otherwise “CANCEL”.
Select Exit to exit from this screen and return to the Setup Configuration”” screen
(see paragraph 4.2.6 “Setup: Configuration”):
4.2.6.2 Setup configuration: Light
Access sequence: RUN>SETTINGS>SETUP>CONFIGURATION>LIGHT
This is used to change the light setting
Select On” to switch on the target illumination;
Select Off” to switch off the target illumination;
Select: Injectionto switch on the target illumination during injection
only; Tap “SAVE” to store the modification, or “CANCEL”.
Select Exit to exit from this screen and return to the Setup Configuration” screen
(see paragraph 4.2.6 “Setup: Configuration“).
4.2.6.3 Setup configuration: Run
Access sequence: RUN>SETTINGS>SETUP>CONFIGURATION>RUN
Use this to change the following operational parameter:
Tray Behaviour: to set tray position during automatic run;
Tray
Behaviour
Closed: the tray remains closed during run
Open: the tray remains open during run
Power-on Restart: sets autosampler restart after a power failure. It should be
set to “ON” only if your GC/analyzer and your data acquisition system are also
able to restart automatically;