HT2800T_en_c.doc Page 126 of 275
Power on-
Restart
On: If the power is lost during a run, when the power comes back on, the
autosampler will restart the injection that it was carrying out at the moment of the
power failure
Off: If the power is lost during a run, when the power comes back on, the
autosampler will return to the home screen (see paragraph 7 . 1 “
screen”), without completing the run.
•
Injection mode: only for liquid and Headspace mode. It sets the predefined injection
mode. This parameter is editable only if the system is configured for two injectors. This
setting is always used for single injection runs, while during an automatic run it is only
used if the injection mode in the injection sequence is set to “Default” (otherwise it will
take the mode specified in the injection sequence). See paragraph 6. 1.2” Setting
sequence step parameters “.
Injection
Mode
Front: the autosampler injects only in the first injector
Rear: the autosampler injects only in the second injector
Confirmation: the autosampler injects the same sample in both injectors
(see “Confirmation mode” explanation in paragraph 1 1 ”Appendix A –
Glossary”).
High Throughput: It injects different sampler alternatively in front and rear
injector (see paragraph 6 . 2 . 2.1” High throughput mode“). This modality
is available only in liquid mode.
•
Desorption mode: only for SPME mode. It sets the predefined desorption mode. This
parameter is editable only if the system is configured for two injectors. This setting is
always used for single injection runs, while during an automatic run it is only used if the
desorption mode in the injection sequence is set to “Default” (otherwise it will take the
mode specified in the injection sequence). See paragraph 6 . 1. 2”Setting sequence
step parameters “.
Desorption
Mode
Front: the desorption is performed in the first injector
Rear: the desorption is performed in the second injector
•
Injection Synchro: only for liquid and Headspace mode. It sets the synchronization
mode between the autosampler and the analyzer or other instruments during the
execution of one or more injections.